Data Foundry

Medical device recalls 2016

1) The HomeChoice automated peritoneal dialysis system — Class II medical device recall Z-2609-2016

Terminated recall by Baxter Healthcare Corp., reported 2016-08-31, distributed nationwide. In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to emp

Recall numberZ-2609-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBaxter Healthcare Corp., Deerfield, IL, United States
Recall initiated2013-10-07
Classified2016-08-22
Reported by FDA2016-08-31
Terminated2016-12-23
DistributedNationwide (US)
Countries outside the USAU, CA, JP, NZ
Reason classessoftware

Product

1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2) The HomeChoice PRO automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. The HomeChoice/HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Reason for recall

In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial Drain. This is to mitigate against Unintended Increased Intraperitoneal Volume (IIVP). This can cause serious problems in patients with unrelated a co-morbid condition of ascites.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2609-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.