t:slim X2 Insulin Pump with Interoperable Technology — Class II medical device recall Z-2621-2025

Ongoing recall by Tandem Diabetes Care, Inc., reported 2025-10-01. an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing iss

Recall numberZ-2621-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmTandem Diabetes Care, Inc., San Diego, CA, United States
Recall initiated2025-07-03
Classified2025-09-24
Reported by FDA2025-10-01
Countries outside the USIL
Quantity19
UPC / GTIN / UDI-DI00389152470641

Product

t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)

Reason for recall

an app defect that occurs when the phone is set to a right-to-left language, which causes app-pump pairing issues and graphical defects, which can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.