Data Foundry

Medical device recalls 2025

Allura CV20 — Class II medical device recall Z-2635-2025

Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2025-10-01, distributed nationwide. Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery,

Recall numberZ-2635-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, N/A, Netherlands
Recall initiated2025-09-03
Classified2025-09-25
Reported by FDA2025-10-01
DistributedNationwide (US)
Countries outside the USAR, AT, AU, BD, BE, BG, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, HN, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KR, LB, LK, LT, MA, MT, MX, MY, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TW, UA, UY, VE, VN, ZA
Quantity46 units
Serial numbers100, 109, 110, 112, 113, 115, 116, 117, 124, 130, 133, 134, 135, 136, 138, 150, 152, 153, 155, 158, 161, 162, 163, 164, 165 and 7 more

Product

Allura CV20; Product Code: 722031; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Reason for recall

Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2635-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.