Data Foundry

Medical device recalls 2025

ZAP-X Radiosurgery System — Class II medical device recall Z-2639-2025

Ongoing recall by Zap Surgical Systems, reported 2025-10-08, distributed nationwide. If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degre

Recall numberZ-2639-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmZap Surgical Systems, San Carlos, CA, United States
Recall initiated2025-07-25
Classified2025-09-26
Reported by FDA2025-10-08
DistributedNationwide (US)
Countries outside the USDE, ES, FR, KR, PL, PY
Quantity17
Reason classessoftware
UPC / GTIN / UDI-DI00860183001504
Serial numbersZUC17153, ZUC20156, ZUC20157, ZUC20159, ZUC21162, ZUC21164, ZUC22166, ZUC22169, ZUC22170, ZUC22171, ZUC22173, ZUC22174, ZUC23176, ZUC23177, ZUC24179, ZUC24180, ZUC24181
Model numbers300150

Product

ZAP-X Radiosurgery System, Model: 300150

Reason for recall

If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.