3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack — Class II medical device recall Z-2640-2023
Ongoing recall by 3M Company - Health Care Business, reported 2023-10-04, distributed nationwide. Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Pl
| Recall number | Z-2640-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | 3M Company - Health Care Business, Saint Paul, MN, United States |
| Recall initiated | 2023-08-22 |
| Classified | 2023-09-26 |
| Reported by FDA | 2023-10-04 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, AU, BE, CA, CN, CO, ES, HK, IT, JP, KR, PE, PL, TW |
| Quantity | 77,280 devices |
| Reason classes | packaging |
| UPC / GTIN / UDI-DI | 50707387787283 |
| Lot numbers | 33P4ND, 33P4YN, 41482VF |
| Model numbers | 41482VF |
| Expiration dates | 2025-01-02 |
Product
3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air removal steam sterilization processes
Reason for recall
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.