Data Foundry

Medical device recalls 2014

ADVIA Chemistry A1c_3 CAL — Class II medical device recall Z-2643-2014

Terminated recall by Siemens Healthcare Diagnostics, reported 2014-09-24, distributed nationwide. The ADVIA Chemistry Systems A1c_3 method may exhibit a percent bias of -9% to +11% for hemoglobin (Hb)A1c samp

Recall numberZ-2643-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Tarrytown, NY, United States
Recall initiated2014-08-06
Classified2014-09-15
Reported by FDA2014-09-24
Terminated2017-02-28
DistributedNationwide (US)
Countries outside the USAR, AT, BR, CA, CH, CN, CO, CY, DE, DK, EE, ES, FI, FR, GR, GT, HR, ID, IN, IT, KR, LT, MX, PE, PL, PY, RO, RS, SG, SK, TH, UY, VN
Lot numbers3HD044
Expiration dates2015-11-01, 2016-01-01

Product

ADVIA Chemistry A1c_3 CAL (Calibrator) for use on the ADVIA Chemistry Systems, REF/Catalog Number/Siemens Material Number (SMN) 10491408, IVD ADVIA Chemistry A1c_3 Calibrator Information: For in vitro diagnostic use in the calibration of A1c_3 (A1c_3 and A1c_3M) and Total Hemoglobin_3 (tHb_3 and tHb_3M) methods on ADVIA Chemistry systems. The ADVIA Chemistry systems are automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine in random access and batch modes for in vitro diagnostic use.

Reason for recall

The ADVIA Chemistry Systems A1c_3 method may exhibit a percent bias of -9% to +11% for hemoglobin (Hb)A1c samples when using A1c_3 Calibrator lots 3HD044 and 3LD068. Depending on quality control limits, this issue may not have been detected. NOTE: The percent bias range is a percentage of the HbA1c result. It is NOT absolute HbA1c units.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2643-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.