Data Foundry

Medical device recalls 2016

DBS Pocket Adaptor — Class II medical device recall Z-2645-2016

Terminated recall by Medtronic Neuromodulation, reported 2016-08-31. Due to the potential for pocket adaptor conductor wire fractures involving DBS Pocket Adaptor Model 64001 and

Recall numberZ-2645-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedtronic Neuromodulation, Minneapolis, MN, United States
Recall initiated2016-06-23
Classified2016-08-24
Reported by FDA2016-08-31
Terminated2017-09-21
Reason classeslabeling, device malfunction
Model numbers64001, 64002

Product

DBS Pocket Adaptor, Model 64001 and Model 64002. The Medtronic Models 64001 (1x4) and 64002 (2x4) Pocket Adaptors can be used as a part of a neurostimulation system for deep brain stimulation. The pocket adaptor is intended to be implanted with the new replacement neurostimulator in the same pocket used for the explanted neurostimulator. Implanting in the same neurostimulator pocket allows for a single-incision procedure.

Reason for recall

Due to the potential for pocket adaptor conductor wire fractures involving DBS Pocket Adaptor Model 64001 and Model 64002, a safety notification was sent to healthcare professionals to reinforce device specific labeling for the handling of pocket adaptors and system integrity checking during implant procedures.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2645-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.