Data Foundry

Medical device recalls 2014

Coulter DxH Diluent — Class II medical device recall Z-2651-2014

Terminated recall by Beckman Coulter Inc., reported 2014-09-24, distributed nationwide. Beckman Coulter is recalling Coulter DxH Diluent because it may be associated with elevated platelet (PLT) bac

Recall numberZ-2651-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc., Brea, CA, United States
Recall initiated2014-08-25
Classified2014-09-16
Reported by FDA2014-09-24
Terminated2015-02-25
DistributedNationwide (US)
Countries outside the USCA, CL, MX, PA, TH
Lot numbers3510470, 3510480, 3510490, 3510500, 3510510, 3510520, 3510530, 3510540, 3510550, 3510570, 3510580, 3510590, 3510600, 3510610, 3510620, 3510630, 3510690
Model numbers628017

Product

Coulter DxH Diluent, Catalog No. 628017 Blood cell analysis comprises diluting a whole-blood sample with a solution that functions as a diluent. The diluent provides the ability to analyze portions of the diluted blood sample for different blood cell types and the measurement of hemoglobin.

Reason for recall

Beckman Coulter is recalling Coulter DxH Diluent because it may be associated with elevated platelet (PLT) background counts between 3-7 x 10(3) cells/uL. This will result in failed Daily Checks on the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2651-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.