Data Foundry

Medical device recalls 2014

VERO Linear Accelerator System — Class II medical device recall Z-2653-2014

Terminated recall by MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK, reported 2014-09-24, distributed in FL, NY, TX. Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would err

Recall numberZ-2653-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK, HIROSHIMA, N/A, Japan
Recall initiated2014-07-05
Classified2014-09-16
Reported by FDA2014-09-24
Terminated2015-04-01
DistributedFL, NY, TX
Quantity3 units
Reason classessoftware
Serial numbers201902, 203901, 203919
Model numbersMHI-TM2000

Product

VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Reason for recall

Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become Treated (completed) status on the exact day when such change is made, to Untreated status. Furthermore, such treatment would be cloned and mistakenly added to the schedule as Untreated treatment

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2653-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.