Data Foundry

Medical device recalls 2023

Veran : Percutaneous Always on Track Models: INS-5040 Always-On… — Class II medical device recall Z-2663-2023

Ongoing recall by Olympus Corporation of the Americas, reported 2023-10-11, distributed nationwide. Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Lap

Recall numberZ-2663-2023
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmOlympus Corporation of the Americas, Center Valley, PA, United States
Recall initiated2023-08-10
Classified2023-09-29
Reported by FDA2023-10-11
DistributedNationwide (US)
Quantity522 units
UPC / GTIN / UDI-DI00815686020217, 00815686020224, 00815686020248, 00815686020255, 00815686020293, 00815686020309, 00815686020330, 00815686020354, 00815686020361, 00815686020699, 00815686020705
Model numbersINS-5040

Product

Veran : Percutaneous Always on Track Models: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga INS-5039 Always-On vTrack Universal Tracker, SM/CL, 12-18ga INS-5036 "22 UTW 125mm Always-On Tip Tracked Biopsy Needle, Trocar point" INS-5034 17 UTW Lg 123mm Always-On Tip Tracked Biopsy Needle, Chiba point 00815686020354 INS-5032 "19 UTW 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" INS-5029 "19 UTW 105 mm Always-On Tip Tracked biopsy needle, Chiba oint" INS-5028 19 UTW 155 mm Always-On Tip Tracked biopsy needle, Chiba point INS-5024 "17ga 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" INS-5023 17ga 100mm Always-On Tip Tracked Biopsy Needle, Chiba point INS-5017 17ga x 155mm Tip Tracked Bx Needle INS-5016 17ga x 155mm Tip Tracked Bx Needle

Reason for recall

Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment