Veran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga… — Class II medical device recall Z-2664-2023
Ongoing recall by Olympus Corporation of the Americas, reported 2023-10-11, distributed nationwide. Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Lap
| Recall number | Z-2664-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Olympus Corporation of the Americas, Center Valley, PA, United States |
| Recall initiated | 2023-08-10 |
| Classified | 2023-09-29 |
| Reported by FDA | 2023-10-11 |
| Distributed | Nationwide (US) |
| Quantity | 60,102 units |
| UPC / GTIN / UDI-DI | 00815686020392, 00815686020408, 00815686020439, 00815686020446, 00815686020477, 00815686020538, 00815686020576, 00815686020590, 00815686020613 |
| Model numbers | INS-0392 |
Product
Veran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx INS-0372 AOTT Serrated Forceps, 1.8mm OD, 5/Bx INS-0362 AOTT Oval Cup Forceps, 1.8mm OD, 5/Bx INS-0352 "Always-On Tip Tracked (AOTT) Brush, 15mm L, 1.8mm OD, 5/Bx" INS-0322 View Peripheral Catheter, 3.2mm OD, 2.0mm WC INS-0305 Always-On Tip Tracked Guidewire, 1.0mm, 950mm INS-0304 Always-On Tip Tracked Guidewire, 1.0mm, 735mm INS-0050 vPad Patient Tracker, 5/Bx INS-0048 vPad Connector Cable
Reason for recall
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment