Siemens CentraLink Data Management System — Class II medical device recall Z-2665-2014
Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2014-09-24, distributed nationwide. Siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink" Data Management
| Recall number | Z-2665-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Healthcare Diagnostics, Inc., Newark, DE, United States |
| Recall initiated | 2014-08-19 |
| Classified | 2014-09-18 |
| Reported by FDA | 2014-09-24 |
| Terminated | 2017-05-12 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BD, BE, BR, CA, CH, CL, CN, CO, DE, DK, EG, ES, FI, FR, GB, GR, HU, IE, IL, IN, IT, KR, LB, LT, MX, MY, NO, NZ, PE, PK, PT, QA, RU, SA, SE, SG, SI, SK, TH, TW, UY, ZA |
| Quantity | 858 |
| Reason classes | software |
| Model numbers | 10810210, 10811596, 10814296, 10814298, 10814877, 10815474 |
Product
Siemens CentraLink Data Management System, software version 14. The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab.
Reason for recall
Siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink" Data Management System, software version 14. The issue may cause the software to stop executing commands, including uploading validated results to the LIS. The issue is related to an internal software timer that overflows after 24 days. Automated rules triggered by task schedulers or by actions in the CentraLink user interface may also be affected. As each customer configuration is unique, in addition to the delay in uploading of validated results, the specific consequences of discontinued command execution will vary.