Data Foundry

Medical device recalls 2014

Siemens CentraLink Data Management System — Class II medical device recall Z-2665-2014

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2014-09-24, distributed nationwide. Siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink" Data Management

Recall numberZ-2665-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Newark, DE, United States
Recall initiated2014-08-19
Classified2014-09-18
Reported by FDA2014-09-24
Terminated2017-05-12
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BD, BE, BR, CA, CH, CL, CN, CO, DE, DK, EG, ES, FI, FR, GB, GR, HU, IE, IL, IN, IT, KR, LB, LT, MX, MY, NO, NZ, PE, PK, PT, QA, RU, SA, SE, SG, SI, SK, TH, TW, UY, ZA
Quantity858
Reason classessoftware
Model numbers10810210, 10811596, 10814296, 10814298, 10814877, 10815474

Product

Siemens CentraLink Data Management System, software version 14. The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab.

Reason for recall

Siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink" Data Management System, software version 14. The issue may cause the software to stop executing commands, including uploading validated results to the LIS. The issue is related to an internal software timer that overflows after 24 days. Automated rules triggered by task schedulers or by actions in the CentraLink user interface may also be affected. As each customer configuration is unique, in addition to the delay in uploading of validated results, the specific consequences of discontinued command execution will vary.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2665-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.