Data Foundry

Medical device recalls 2014

The Ysio Systems with software version VC10 The Ysio enables… — Class II medical device recall Z-2680-2014

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2014-10-01, distributed nationwide. Siemens discovered that an unlikely error may occur on the Ysio system with fixed detector in the wall stand.

Recall numberZ-2680-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2014-08-27
Classified2014-09-23
Reported by FDA2014-10-01
Terminated2015-03-30
DistributedNationwide (US)
Quantity62
Model numbers10281013, 10281163

Product

The Ysio Systems with software version VC10 The Ysio enables radiographic and tomographic exposures of the whole Body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.

Reason for recall

Siemens discovered that an unlikely error may occur on the Ysio system with fixed detector in the wall stand. This error may result in line artifacts in the image. If these artifacts appear in the region of interest, the examination may need to be repeated.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2680-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.