The Verigene Clostridium difficile Nucleic Acid Test — Class II medical device recall Z-2682-2014
Terminated recall by Nanosphere, Inc., reported 2014-10-01, distributed in AZ, CA, FL, GA, IA, IL, IN, KY…. Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures p
| Recall number | Z-2682-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Nanosphere, Inc., Northbrook, IL, United States |
| Recall initiated | 2013-10-10 |
| Classified | 2014-09-23 |
| Reported by FDA | 2014-10-01 |
| Terminated | 2015-06-16 |
| Distributed | AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI |
| Countries outside the US | BG, DE, ES, IT, JP, KR |
| Quantity | 80 units |
| Lot numbers | 100313022B |
| Expiration dates | 2014-04-01 |
Product
The Verigene Clostridium difficile Nucleic Acid Test (CDF) is a qualitative, multiplexed in vitro diagnostic test for the rapid detection of toxin A (tcdA), toxin B (tcdB), and tcdC gene sequences of toxigenic strains Clostridium difficile and for presumptive identification of PCR ribotype 027 strains from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI). Presumptive identification of the PCR ribotype 027 strain of C. difficile is by detection of the binary toxin (cdt) gene sequence and the single base pair deletion at nucleotide 117 in the tcdC gene. The tcdC gene encodes for a negative regulator in C. difficile toxin production. The test is performed on the Verigene System and utilizes automated specimen preparation and polymerase chain reaction (PCR) amplification, combined with a nanoparticle-based array hybridization assay to detect the toxin gene sequences associated with toxin-producing C. difficile.
Reason for recall
Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.
Machine access