Nicolet¿ Elite¿ 3 MHz OB Probe — Class II medical device recall Z-2728-2017
Terminated recall by Natus Neurology Inc, reported 2017-07-19, distributed in AK, AL, AR, AZ, CA, CO, CT, FL…. Some of the Elite 2 and 3 Obstetric probes and 5 MHz Vascular Probes may exhibit excessive background noise th
| Recall number | Z-2728-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Natus Neurology Inc, Middleton, WI, United States |
| Recall initiated | 2017-06-09 |
| Classified | 2017-07-13 |
| Reported by FDA | 2017-07-19 |
| Terminated | 2020-07-22 |
| Distributed | AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV |
| Countries outside the US | AU, CA, CH, CL, DE, GB, IN, JO, JP, KW, NL, PH, QA, SA, SE, SG, TH |
| Quantity | 2,335 units |
| Serial numbers | CAQ0100, CAQ0101, CAQ0102, CAQ0103, CAQ0104, CAQ0105, CAQ0106, CAQ0107, CAQ0108, CAQ0109, CAQ0110, CAQ0111, CAQ0112, CAQ0113, CAQ0114, CAQ0115, CAQ0116, CAQ0117, CAQ0118, CAQ0119, CAQ0120, CAQ0121, CAQ0122, CAQ0123, CAQ0124 and 475 more |
| Model numbers | N300, X1L001 |
Product
Nicolet¿ Elite¿ 3 MHz OB Probe, Model #: X1L001, Catalog Code: N300. Intended for non-invasive support, i.e. early detection of fetal life, monitor blood flow before and after surgery, etc.
Reason for recall
Some of the Elite 2 and 3 Obstetric probes and 5 MHz Vascular Probes may exhibit excessive background noise that may mask the users ability to clearly hear heart rate and/or blood flow.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2728-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.