Centricity Enterprise Archive is a software product for receiving — Class II medical device recall Z-2739-2016
Terminated recall by Ge Healthcare It, reported 2016-09-14, distributed nationwide. There is a potential for data loss associated with empty path names resulting from HL-7 updates when the cache
| Recall number | Z-2739-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ge Healthcare It, Barrington, IL, United States |
| Recall initiated | 2012-11-16 |
| Classified | 2016-09-06 |
| Reported by FDA | 2016-09-14 |
| Terminated | 2017-01-11 |
| Distributed | Nationwide (US) |
| Quantity | 1,379 units |
Product
Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. Qualified system administrators install, monitor and maintain the system. DICOM devices (e.g. modalities, workstations) communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction other systems have with the Centricity Enterprise Archive. Centricity Enterprise Archive is not intended for use in patient vicinity. The intended users for Centricity Enterprise Users service personnel and System administrators, and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-I profiles (published by IHE).
Reason for recall
There is a potential for data loss associated with empty path names resulting from HL-7 updates when the cache disk is almost full. In an unlikely event that the cache disk is almost full (<100MB) and HL-7 update events are received, the update action fails and an empty path name is written to the EA database. As a result the instance is not deleted from the storage device but no longer referenced. When trying to retrieve the instance an error is returned.