Data Foundry

Medical device recalls 2020

DCA Vantage Handheld Barcode Scanner — Class II medical device recall Z-2747-2020

Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2020-08-12, distributed nationwide. If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integ

Recall numberZ-2747-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Inc., Newark, DE, United States
Recall initiated2020-06-30
Classified2020-07-31
Reported by FDA2020-08-12
Terminated2020-11-17
DistributedNationwide (US)
Quantity1,800 units
Model numbersDS4308, HC0015BZZWW

Product

DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US

Reason for recall

If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2747-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.