Data Foundry

Medical device recalls 2018

CDI Blood parameter monitoring system 500 H/SAT — Class II medical device recall Z-2750-2018

Terminated recall by Terumo Cardiovascular Systems Corporation, reported 2018-08-22, distributed nationwide. H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an incre

Recall numberZ-2750-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmTerumo Cardiovascular Systems Corporation, Ann Arbor, MI, United States
Recall initiated2018-06-26
Classified2018-08-10
Reported by FDA2018-08-22
Terminated2020-04-17
DistributedNationwide (US)
Countries outside the USCA, JP, MX
Quantity128 in
Model numbers145883

Product

CDI Blood parameter monitoring system 500 H/SAT, Part No. 145883 Product Usage: The CDI System 500 provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature. In addition, calculated values of base excess, bicarbonate, oxygen saturation, and oxygen consumption may also be provided. These parameters are displayed at either actual temperature or adjusted to 37¿C. For documentation purposes, the system 500s integral printer provides a hard copy of displayed parameters.

Reason for recall

H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automated self-test when the device is powered on. This will not result in device inaccuracy, only Color Chip Test Failure which disables the H/Sat module.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2750-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.