Monitor with one Blood Parameter Module and one HSat Probe — Class II medical device recall Z-2751-2018
Terminated recall by Terumo Cardiovascular Systems Corporation, reported 2018-08-22, distributed nationwide. H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an incre
| Recall number | Z-2751-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Terumo Cardiovascular Systems Corporation, Ann Arbor, MI, United States |
| Recall initiated | 2018-06-26 |
| Classified | 2018-08-10 |
| Reported by FDA | 2018-08-22 |
| Terminated | 2020-04-17 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA, JP, MX |
| Quantity | 128 in |
Product
Monitor with one Blood Parameter Module and one HSat Probe; 500AHCT Monitor with two Blood Parameter modules and one HSat Probe; 500AVHCT Product Usage: The CDI System 500 provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature. In addition, calculated values of base excess, bicarbonate, oxygen saturation, and oxygen consumption may also be provided. These parameters are displayed at either actual temperature or adjusted to 37¿C. For documentation purposes, the system 500s integral printer provides a hard copy of displayed parameters.
Reason for recall
H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automated self-test when the device is powered on. This will not result in device inaccuracy, only Color Chip Test Failure which disables the H/Sat module.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2751-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.