FARADRIVE Steerable Sheath Clear IDE — Class II medical device recall Z-2754-2026
Ongoing recall by Boston Scientific Corporation, reported 2026-07-29, distributed nationwide. Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are assoc
| Recall number | Z-2754-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Boston Scientific Corporation, Saint Paul, MN, United States |
| Recall initiated | 2026-06-11 |
| Classified | 2026-07-17 |
| Reported by FDA | 2026-07-29 |
| Distributed | Nationwide (US) |
| Quantity | 400 units |
| UPC / GTIN / UDI-DI | 00810087181062 |
| Lot numbers | CL12910, CL12978, CL15153 |
Product
FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01
Reason for recall
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2754-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.