Data Foundry

Medical device recalls 2017

Medtronic Visualase — Class II medical device recall Z-2755-2017

Terminated recall by Medtronic Navigation, Inc., reported 2017-08-02, distributed in CA, FL, GA, IL, MN, NY, TX. Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not

Recall numberZ-2755-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmMedtronic Navigation, Inc., Louisville, CO, United States
Recall initiated2017-06-07
Classified2017-07-21
Reported by FDA2017-08-02
Terminated2018-06-29
DistributedCA, FL, GA, IL, MN, NY, TX
Quantity214 units
Reason classespackaging
Lot numbers12473141, 211158059, 211158151, 211158152, 211181212, 211181547, 211184941, 211555011, 211555017, 211560973, 211594328, 211599898, 211628257, 211908695, 211916062, 211926889, 211950736, 211959879, 211977586, 211987873, 212127654, 212141557, 212148368, 212434500, 212438405, 212503634, 212504463, 212549456, 212553428, 212632884, 212647830, 212701679, 212707035, 212745660, 212796559, 212912501
Model numbers9735561

Product

Medtronic Visualase(R) Cooled Laser Applicator System (VCLAS) 15MM TIP, REF 9735561, (1) Laser Diffusing Fiber - 600um Core, 15mm Tip, (1) Cooling Catheter with 3M(TM) SteriStrip(TM), (1) Pump Tubing Set, (1) Drainage Bag. Medtronic Navigation, Inc. Louisville, CO 80027

Reason for recall

Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling system. When the TBA is unable to seal around the optical fiber, low retention force to the optical fiber may occur, causing saline to leak out of the TBA.