Data Foundry

Medical device recalls 2026

FARADRIVE STEERABLE SHEATH CLEAR — Class II medical device recall Z-2755-2026

Ongoing recall by Boston Scientific Corporation, reported 2026-07-29, distributed nationwide. Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are assoc

Recall numberZ-2755-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Corporation, Saint Paul, MN, United States
Recall initiated2026-06-11
Classified2026-07-17
Reported by FDA2026-07-29
DistributedNationwide (US)
Quantity74,760 units
UPC / GTIN / UDI-DI00810087180072
Lot numbersCL12697, CL12698, CL12701, CL12702, CL12824, CL12832, CL12833, CL12861, CL12888, CL12889, CL12912, CL12913, CL12916, CL12918, CL12968, CL12975, CL12977, CL13003, CL13016, CL13019, CL13045, CL13047, CL13059, CL13078, CL13079, CL13101, CL13102, CL13120, CL13121, CL13140, CL13141, CL13158, CL13159, CL13195, CL13211, CL13218, CL13232, CL13233, CL13259, CL13290 and 67 more

Product

FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21M402, IFU Reference Number LBL1062

Reason for recall

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2755-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.