Aortic Arch Cannula and Coronary Artery Perfusion Cannula with Ballon — Class II medical device recall Z-2756-2016

Terminated recall by Sorin Group USA, Inc., reported 2016-09-14, distributed in IL, MO, TX. Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Ca

Recall numberZ-2756-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSorin Group USA, Inc., Arvada, CO, United States
Recall initiated2016-08-12
Classified2016-09-08
Reported by FDA2016-09-14
Terminated2017-03-27
DistributedIL, MO, TX
Countries outside the USCA, IT
Quantity200 units
Lot numbers1526500032, 1529300026, 1602600042
Model numbersNA-55X6

Product

Aortic Arch Cannula and Coronary Artery Perfusion Cannula with Ballon (part number NA-55X6). The Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect. The Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip

Reason for recall

Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Cannula With Ballon because the tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.