Aortic Arch Cannula and Coronary Artery Perfusion Cannula with Ballon — Class II medical device recall Z-2756-2016
Terminated recall by Sorin Group USA, Inc., reported 2016-09-14, distributed in IL, MO, TX. Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Ca
| Recall number | Z-2756-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Sorin Group USA, Inc., Arvada, CO, United States |
| Recall initiated | 2016-08-12 |
| Classified | 2016-09-08 |
| Reported by FDA | 2016-09-14 |
| Terminated | 2017-03-27 |
| Distributed | IL, MO, TX |
| Countries outside the US | CA, IT |
| Quantity | 200 units |
| Lot numbers | 1526500032, 1529300026, 1602600042 |
| Model numbers | NA-55X6 |
Product
Aortic Arch Cannula and Coronary Artery Perfusion Cannula with Ballon (part number NA-55X6). The Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect. The Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip
Reason for recall
Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Cannula With Ballon because the tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.