Power F2 Powered Fastening System — Class II medical device recall Z-2760-2026
Ongoing recall by Ferno-Washington Inc, reported 2026-07-29, distributed nationwide. Product may experience accelerated roller wear and potential deformation due to interference between the rolle
| Recall number | Z-2760-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ferno-Washington Inc, Wilmington, OH, United States |
| Recall initiated | 2026-06-19 |
| Classified | 2026-07-17 |
| Reported by FDA | 2026-07-29 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA, DE, DK, GB, NL, NO, NZ, PL |
| Quantity | 278 |
| UPC / GTIN / UDI-DI | 00190790006570 |
| Serial numbers | 25C227687, 25C227688, 25C227689, 25C227690, 25C227691, 25C227692, 25C227693, 25C227694, 25C227695, 25C227696, 25C227697, 25C227698, 25C227699, 25C227700, 25C227701, 25C227702, 25C227703, 25C227704, 25C227707, 25C227708, 25C227709, 25C227710, 25C227711, 25C228204, 25C228205 and 252 more |
| Model numbers | 0567091 |
Product
Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091
Reason for recall
Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2760-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.