Data Foundry

Medical device recalls 2020

Captivator Polypectomy Snare — Class II medical device recall Z-2761-2020

Ongoing recall by Boston Scientific Corporation, reported 2020-08-12. BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences re

Recall numberZ-2761-2020
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Corporation, Marlborough, MA, United States
Recall initiated2020-07-06
Classified2020-08-05
Reported by FDA2020-08-12
UPC / GTIN / UDI-DI08714729019336
Lot numbers25158564, 25245733, 25245735, 25330549, 25403001, 25439073, 25513671
Expiration dates2023-02-04

Product

Captivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loop.

Reason for recall

BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2761-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.