Captivator Polypectomy Snare — Class II medical device recall Z-2762-2020
Ongoing recall by Boston Scientific Corporation, reported 2020-08-12. BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences re
| Recall number | Z-2762-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Boston Scientific Corporation, Marlborough, MA, United States |
| Recall initiated | 2020-07-06 |
| Classified | 2020-08-05 |
| Reported by FDA | 2020-08-12 |
| UPC / GTIN / UDI-DI | 08714729019350 |
| Lot numbers | 25077093, 25164173, 25185062, 25187234, 25226123, 25255681, 25336754, 25403003, 25508379, 25524901 |
| Expiration dates | 2023-01-20 |
Product
Captivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loop.
Reason for recall
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2762-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.