Data Foundry

Medical device recalls 2015

Pointe Scientific autoHDL Reagent H7545 H7545-R1 in a bulk cube… — Class III medical device recall Z-2764-2015

Terminated recall by Pointe Scientific, Inc., reported 2015-09-23, distributed in CA, CT, FL, MI, PR, TX. Crystals may be observed in the R1 reagent. Crystals may be a result of the reagent being frozen during transp

Recall numberZ-2764-2015
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPointe Scientific, Inc., Canton, MI, United States
Recall initiated2015-03-13
Classified2015-09-16
Reported by FDA2015-09-23
Terminated2015-10-29
DistributedCA, CT, FL, MI, PR, TX
Countries outside the USCA
Lot numbers430402
Model numbers12-H7545-162, H7545-1000, H7545-R1, HDL600

Product

Pointe Scientific autoHDL Reagent H7545 H7545-R1 in a bulk cube container or HDPE plastic bottle H7545-1000 kit includes 1x750 ml of R1 and 1x250 ml of R2 12-H7545-162 kit includes 3x40 ml of R1 and 3x14 ml of R2 HDL600 kit includes 3x40 ml of R1 and 3x14 ml of R2 Kits: HDPE plastic bottle, PP plastic caps. Bulk: PE plastic cube containers, PP plastic caps. Quantitative determination of high density lipoprotein cholesterol in human serum or plasma. In vitro diagnostic use only.

Reason for recall

Crystals may be observed in the R1 reagent. Crystals may be a result of the reagent being frozen during transport. The crystals do not impact product performance.