Data Foundry

Medical device recalls 2015

Boston Scientific — Class II medical device recall Z-2765-2015

Terminated recall by Boston Scientific Corporation, reported 2015-09-23, distributed nationwide. Cross-threading of the hemostasis valve may occur if it is tightened with the dilator in place, possibly preve

Recall numberZ-2765-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Corporation, Maple Grove, MN, United States
Recall initiated2015-08-25
Classified2015-09-15
Reported by FDA2015-09-23
Terminated2017-04-04
DistributedNationwide (US)
Countries outside the USAE, AR, AT, AU, BE, CA, CH, CL, CN, CO, CZ, DE, DK, EC, EG, ES, FI, FR, GB, GR, HK, HU, ID, IE, IL, IR, IT, KR, KW, LB, MX, MY, NL, NZ, OM, PA, PH, PL, PT, RO, RS, RU, SA, SE, SG, SI, TH, TN, TR, TW, ZA
Quantity29,169 units
Reason classespackaging

Product

Boston Scientific, WATCHMAN Left Atrial Appendage Closure Device with Delivery System, 21mm x 12F ( M635WU21060, WU21060, 24mm x 12F(M635WU24060, WU2406) , 27mm x 12F(M635WU27060, WU2706) , 30mm x 12F(M635WU30060, WU3006) , 33mm x 12F( M635WU33060, WU3306). For cardiac use.

Reason for recall

Cross-threading of the hemostasis valve may occur if it is tightened with the dilator in place, possibly preventing subsequent sealing of the valve when desired. This Field Correction reinforces the Directions for Use (DFU) and gives further guidance on correct use of the hemostasis valve to avoid cross-threading and to securely seal the valve, minimizing potential for undesired blood leakage.