Data Foundry

Medical device recalls 2019

Accu-Chek Aviva II Meter — Class II medical device recall Z-2765-2019

Terminated recall by Roche Diabetes Care, Inc., reported 2019-10-09, distributed nationwide. Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or me

Recall numberZ-2765-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoche Diabetes Care, Inc., Indianapolis, IN, United States
Recall initiated2017-06-12
Classified2019-09-30
Reported by FDA2019-10-09
Terminated2020-10-30
DistributedNationwide (US)
Quantity26,600

Product

Accu-Chek Aviva II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.

Reason for recall

Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2765-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.