Accu-Chek Performa II Meter — Class II medical device recall Z-2766-2019
Terminated recall by Roche Diabetes Care, Inc., reported 2019-10-09, distributed nationwide. Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or me
| Recall number | Z-2766-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Roche Diabetes Care, Inc., Indianapolis, IN, United States |
| Recall initiated | 2017-06-12 |
| Classified | 2019-09-30 |
| Reported by FDA | 2019-10-09 |
| Terminated | 2020-10-30 |
| Distributed | Nationwide (US) |
Product
Accu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.
Reason for recall
Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).