NICO Myriad Console Model number NN-7000 The Myriad System is a… — Class II medical device recall Z-2767-2015
Terminated recall by Nico Corp., reported 2015-09-23, distributed in CA, IN, MD, NC. Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the
| Recall number | Z-2767-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Nico Corp., Indianapolis, IN, United States |
| Recall initiated | 2015-07-29 |
| Classified | 2015-09-15 |
| Reported by FDA | 2015-09-23 |
| Terminated | 2015-09-22 |
| Distributed | CA, IN, MD, NC |
| Quantity | 6 units |
| Reason classes | device malfunction |
| Model numbers | 00231, 00232, 00242, 00245, 00248, 00250, NN-7000 |
Product
NICO Myriad Console Model number NN-7000 The Myriad System is a surgical system specifically designed for morcellation of soft tissue. It primarily consists of a reusable console with foot pedal and a disposable handpiece. For the morcellation and removal of tissue during pelviscopic, laparoscopic, percutaneous and open surgical procedures whenever access to the surgical site is limited.
Reason for recall
Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.