Data Foundry

Medical device recalls 2016

CARTO 3 EP Navigation System and Accessories — Class II medical device recall Z-2776-2016

Terminated recall by Biosense Webster, Inc., reported 2016-09-21. Biosense Webster, Inc. is initiating a voluntary correction to re-install identified CARTO 3 Systems that were

Recall numberZ-2776-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBiosense Webster, Inc., Irwindale, CA, United States
Recall initiated2016-08-15
Classified2016-09-09
Reported by FDA2016-09-21
Terminated2017-01-20
Countries outside the USBR, CN, HK
Quantity338 units

Product

CARTO 3 EP Navigation System and Accessories, Product Codes: FG540000 and FG540000C, Software Versions: v 3.2 , v 4.3 and v 2.3. The CARTO 3 EP Navigation System is intended to be used in catheter-based cardiac electrophysiological (EP) procedures. The CARTO 3 System provides information about the electrical activity of the heart and catheter location during the procedure.

Reason for recall

Biosense Webster, Inc. is initiating a voluntary correction to re-install identified CARTO 3 Systems that were moved and installed in China without testing and approval by Biosense Webster Technical Services, as described in the Instructions for Use (IFU).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2776-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.