CARTO 3 EP Navigation System and Accessories — Class II medical device recall Z-2776-2016
Terminated recall by Biosense Webster, Inc., reported 2016-09-21. Biosense Webster, Inc. is initiating a voluntary correction to re-install identified CARTO 3 Systems that were
| Recall number | Z-2776-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Biosense Webster, Inc., Irwindale, CA, United States |
| Recall initiated | 2016-08-15 |
| Classified | 2016-09-09 |
| Reported by FDA | 2016-09-21 |
| Terminated | 2017-01-20 |
| Countries outside the US | BR, CN, HK |
| Quantity | 338 units |
Product
CARTO 3 EP Navigation System and Accessories, Product Codes: FG540000 and FG540000C, Software Versions: v 3.2 , v 4.3 and v 2.3. The CARTO 3 EP Navigation System is intended to be used in catheter-based cardiac electrophysiological (EP) procedures. The CARTO 3 System provides information about the electrical activity of the heart and catheter location during the procedure.
Reason for recall
Biosense Webster, Inc. is initiating a voluntary correction to re-install identified CARTO 3 Systems that were moved and installed in China without testing and approval by Biosense Webster Technical Services, as described in the Instructions for Use (IFU).