Data Foundry

Medical device recalls 2017

STOCKERT Heater-Cooler System 3T — Class II medical device recall Z-2776-2017

Terminated recall by Sorin Group USA, Inc., reported 2017-08-02, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenato

Recall numberZ-2776-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSorin Group USA, Inc., Arvada, CO, United States
Recall initiated2017-06-30
Classified2017-07-27
Reported by FDA2017-08-02
Terminated2023-10-10
DistributedAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Quantity7,546 units
Serial numbers16S10958-16S16847
Model numbers16-02-85

Product

STOCKERT Heater-Cooler System 3T, 200V/50 Hz/60 Hz

Reason for recall

Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2776-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.