Data Foundry

Medical device recalls 2016

Integra Meshed Dermal Regeneration Template 5 cmx 5 cm — Class II medical device recall Z-2777-2016

Terminated recall by Integra LifeSciences Corporation, reported 2016-09-21, distributed nationwide. Integra's post QA release review of historical product release test results for Meshed IDRT products identifie

Recall numberZ-2777-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmIntegra LifeSciences Corporation, Plainsboro, NJ, United States
Recall initiated2016-03-11
Classified2016-09-09
Reported by FDA2016-09-21
Terminated2016-09-13
DistributedNationwide (US)
Quantity14 packages
Reason classespotency
Lot numbers105A00324750

Product

Integra Meshed Dermal Regeneration Template 5 cmx 5 cm (2 in x 2in) Rx Only Meshed Integra¿ Dermal Regeneration Template, (Integra template) is a bilayer membrane system for skin replacement. The dermal replacement layer is made of a porous matrix of fibers of cross-linked bovine tendon collagen and glycosaminoglycan (chondroitin-6-sulfate) that is manufactured with a controlled porosity and defined degradation rate. The epidermal substitute layer is made of a thin polysiloxane (silicone) layer to control moisture loss from the wound. Integra template is indicated for the postexcisional treatment of life-threatening full-thickness or deep partial-thickness thermal injuries where sufficient autograft is not available at the time of excision or not desirable due to the physiological condition of the patient. Integra template is also indicated for the repair of scar contractures when other therapies have failed or when donor sites for repair are not sufficient or desirable due to the physiological condition of the patient Manufactured by: Integra LifeSciences Corporation 311 Enterprise Drive, Plainsboro, NJ 08536 877-444-1122 USA n 609-936-5400 outside USA 866-800-7742 fax

Reason for recall

Integra's post QA release review of historical product release test results for Meshed IDRT products identified the Peel Strength test average result was incorrectly calculated for a single Lot (Lot 105A00324750).

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2777-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.