Data Foundry

Medical device recalls 2015

Dynamic Locking Screw System — Class II medical device recall Z-2794-2015

Terminated recall by Synthes (USA) Products LLC, reported 2015-09-30, distributed nationwide. DLS pin breakage during planned implant removal, after uneventful and successful healing of the fracture. This

Recall numberZ-2794-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSynthes (USA) Products LLC, West Chester, PA, United States
Recall initiated2015-08-11
Classified2015-09-18
Reported by FDA2015-09-30
Terminated2016-08-18
DistributedNationwide (US)
Quantity40,038
Reason classesdevice malfunction
Lot numbers2748748, 2748939, 2748940, 2748943, 2748945, 2748947, 2748971, 2748972, 2748973, 2748974, 2748975, 2748976, 2748977, 2748978, 2749819, 2749820, 2749821, 2749825, 2749827, 2749829, 2749831, 2751623, 2751624, 2751625, 2751626, 2751632, 2763692, 2763694, 2773533, 2773534, 2773536, 2773539, 2774563, 2774564, 2778498, 2778507, 2778508, 2778519, 2778525, 2778526 and 460 more
Model numbers09.213.022S, 09.213.024S, 09.213.026S, 09.213.028S, 09.213.030S, 09.213.032S, 09.213.034S, 09.213.036S, 09.213.038S, 09.213.040S, 09.213.042S, 09.213.044S, 09.213.046S, 09.213.048S, 09.213.050S, 09.213.052S, 09.213.054S, 09.213.056S, 09.213.058S, 09.213.060S, 09.213.065S, 09.213.070S, 09.223.032S, 09.223.034S, 09.223.036S and 15 more

Product

Dynamic Locking Screw System (DLS) 3.7mm & Dynamic Locking Screw (DLS) 5.0mm; intended for use in long bone fractures, the fixation of osteopenic bone, the fixation of osteotomies, and for the fixation of non-unions and malunions.

Reason for recall

DLS pin breakage during planned implant removal, after uneventful and successful healing of the fracture. This notice to additional consignees identified after recall Z-0390-2014, related to complaints of breakage located at the bottom of the screws and under the locking screw head.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2794-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.