Data Foundry

Medical device recalls 2020

uEXPLORER PET/CT Diagnostic Imaging System — Class II medical device recall Z-2799-2020

Terminated recall by Shanghai United Imaging Healthcare Co., Ltd., reported 2020-08-19, distributed nationwide. 1) 1) Potential intermittent issue may cause image data problems, may lead to artifacts; 2) Potential intermit

Recall numberZ-2799-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmShanghai United Imaging Healthcare Co., Ltd., Shanghai, N/A, China
Recall initiated2020-07-16
Classified2020-08-13
Reported by FDA2020-08-19
Terminated2022-07-07
DistributedNationwide (US)
Quantity1 units
UPC / GTIN / UDI-DI06971576832040
Serial numbersF00002

Product

uEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Reason for recall

1) 1) Potential intermittent issue may cause image data problems, may lead to artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2799-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.