Data Foundry

Medical device recalls 2018

CoLink — Class II medical device recall Z-2880-2018

Terminated recall by In2bones USA, LLC, reported 2018-08-29, distributed nationwide. CoLink(TM) Lapidus Plate +2mm Right and Left, were laser marked incorrectly. In these lots, right plates were

Recall numberZ-2880-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmIn2bones USA, LLC, Memphis, TN, United States
Recall initiated2018-07-16
Classified2018-08-21
Reported by FDA2018-08-29
Terminated2019-02-22
DistributedNationwide (US)
Countries outside the USFR
Quantity48 units
Lot numbers1704075
Model numbersIN2BONES, P40, ST265

Product

CoLink(TM) Lapidus Plate XP, +2 mm, Left, REF P40 ST265, QTY 1, In2Bones Product Usage: The In2Bones USA LLC, CoLink¿ Plating System is indicated for stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction of the small bones in the hand, wrist, foot and ankle in both pediatric and adult patients.

Reason for recall

CoLink(TM) Lapidus Plate +2mm Right and Left, were laser marked incorrectly. In these lots, right plates were laser marked as left plates and the left plates were laser marked as right plates.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2880-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.