BD Alaris System — Class I medical device recall Z-2882-2020
Ongoing recall by CareFusion 303, Inc., reported 2020-09-16, distributed in AK, AL, AR, AS, AZ, CA, CO, CT…. The Alaris PC unit can display incorrect syringe type and/or syringe sizes. This could result in delays in inf
| Recall number | Z-2882-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | CareFusion 303, Inc., San Diego, CA, United States |
| Recall initiated | 2020-08-04 |
| Classified | 2020-09-04 |
| Reported by FDA | 2020-09-16 |
| Distributed | AK, AL, AR, AS, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY |
| Quantity | 180,186 parts |
| Model numbers | 8110, 8120 |
Product
BD Alaris System; Alaris Syringe Module Model 8110; BD Alaris System; Alaris PCA Module Model 8120; BD Alaris System; Syringe/PCA Sizer Sensor Replacement Kit, P/N 122278652
Reason for recall
The Alaris PC unit can display incorrect syringe type and/or syringe sizes. This could result in delays in infusion, under-infusion, or over-infusion.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2882-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.