Data Foundry

Medical device recalls 2016

Integra Ventricular Catheter Accessory Kit — Class II medical device recall Z-2909-2016

Terminated recall by Integra LifeSciences Corp., reported 2016-10-12, distributed nationwide. During final sampling inspection, one of the lots was discovered with a total incomplete packaging seal of the

Recall numberZ-2909-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmIntegra LifeSciences Corp., Plainsboro, NJ, United States
Recall initiated2016-08-23
Classified2016-09-30
Reported by FDA2016-10-12
Terminated2017-03-29
DistributedNationwide (US)
Quantity14
Reason classessterility, packaging
Lot numbers0195786
Model numbers951303
Expiration dates2019-05

Product

Integra Ventricular Catheter Accessory Kit, REF 951303, Sterile EO, Rx only, packaged in a PVC tray, then secondary package in to Tyvek pouch and tertiary packaged in to Tyvek pouch. Sealed pouches are packaged in a cardboard box. F8 35cm Silicone Elastomer Ventricular Catheter for use with CSF external drainage systems

Reason for recall

During final sampling inspection, one of the lots was discovered with a total incomplete packaging seal of the sterile product.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2909-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.