Data Foundry

Medical device recalls 2026

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000 — Class II medical device recall Z-2938-2026

Ongoing recall by Erbe USA Inc, reported 2026-08-19, distributed nationwide. A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit betw

Recall numberZ-2938-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmErbe USA Inc, Marietta, GA, United States
Recall initiated2026-06-26
Classified2026-08-12
Reported by FDA2026-08-19
DistributedNationwide (US)
Countries outside the USMX
Quantity1,126 units
Reason classespackaging, device malfunction
UPC / GTIN / UDI-DI04050147013797
Serial numbers11608268
Model numbers10402-000

Product

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

Reason for recall

A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2938-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.