BD Alaris System PC Unit Model 8100 — Class I medical device recall Z-2939-2020

Ongoing recall by CareFusion 303, Inc., reported 2020-09-16, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially re

Recall numberZ-2939-2020
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCareFusion 303, Inc., San Diego, CA, United States
Recall initiated2020-08-04
Classified2020-09-10
Reported by FDA2020-09-16
DistributedAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Countries outside the USAE, AU, CA, HU, IL, JP, KW, NZ, PH, QA, SA, SG, TR, TW
Quantity248,519
Model numbers8100

Product

BD Alaris System PC Unit Model 8100 (Pump Module) and Pump Module Door Assembly Replacement Kits (P/N 49000239, 49000346, 49000438, 49000439) infusion pump.

Reason for recall

Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious injury or death in high-risk patient populations.