Data Foundry

Medical device recalls 2018

Roche cobas p 612 pre-analytical system — Class II medical device recall Z-2944-2018

Terminated recall by Roche Diagnostics Corporation, reported 2018-09-05, distributed in CA, NY, TX, WA, WI. Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-

Recall numberZ-2944-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoche Diagnostics Corporation, Indianapolis, IN, United States
Recall initiated2018-07-16
Classified2018-08-27
Reported by FDA2018-09-05
Terminated2019-05-16
DistributedCA, NY, TX, WA, WI
Quantity12 units
Model numbers07563116001

Product

Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cobas p 612 pre-analytical system is a computer controlled fully automatic system for sorting of open and closed barcoded and centrifuged and non-centrifuged sample tubes. The cobas p 612 is intended for use with analyzers that perform tests in the area of clinical chemistry, immuno chemistry, coagulation, hematology, urinalysis, nucleic acid testing.

Reason for recall

Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-analytical system (63x) (LCP1) with non-filter tips. The possible presence of biological material on the nozzle may cause contamination resulting in potential false positive results, depending on the sensitivity of the analytical technology.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2944-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.