25-LHP-828 — Class II medical device recall Z-2975-2024

Ongoing recall by PACIFIC LASERTECH, LLC, reported 2024-09-18, distributed nationwide. These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system cl

Recall numberZ-2975-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedFDA mandated
Recalling firmPACIFIC LASERTECH, LLC, San Marcos, CA, United States
Recall initiated2024-03-01
Classified2024-09-10
Reported by FDA2024-09-18
DistributedNationwide (US)
Quantity4

Product

25-LHP-828, HeNe Laser System, 35 mW

Reason for recall

These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.