Data Foundry

Medical device recalls 2026

MiniMed 780G Insulin Pump — Class II medical device recall Z-2982-2026

Ongoing recall by Medtronic MiniMed, Inc., reported 2026-08-26. Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump ca

Recall numberZ-2982-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedtronic MiniMed, Inc., Northridge, CA, United States
Recall initiated2026-07-13
Classified2026-08-18
Reported by FDA2026-08-26
Countries outside the USAT, BE, CH, DE, DK, GB, IE, IT, NL, SE
Quantity257
UPC / GTIN / UDI-DI00763000442187, 00763000442194, 00763000974374, 00763000974381
Serial numbers00763000974374/NG3150737H, 00763000974381/NG2966956H, MMT-1885CLW/00763000442187, MMT-1886CLW/00763000442194, NG2854971H, NG2901117H, NG2970938H, NG2986362H, NG3013390H, NG3057910H, NG3080842H, NG3081463H, NG3081498H, NG3081525H, NG3081894H, NG3081975H, NG3081976H, NG3081982H, NG3082019H, NG3082036H, NG3085044H, NG3088436H, NG3088708H, NG3090073H, NG3090722H and 234 more
Model numbersMMT-1885CLW, MMT-1886CLW

Product

MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW

Reason for recall

Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-2982-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.