Alere iScreen Single Drug Detection — Class II medical device recall Z-3029-2020
Terminated recall by Ameditech Inc, reported 2020-10-07, distributed nationwide. Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all ca
| Recall number | Z-3029-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ameditech Inc, San Diego, CA, United States |
| Recall initiated | 2020-08-18 |
| Classified | 2020-09-30 |
| Reported by FDA | 2020-10-07 |
| Terminated | 2024-09-30 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, DE, ES, FR, NZ, SG |
| Quantity | 303,100 units |
| Reason classes | labeling |
Product
Alere iScreen Single Drug Detection (AMT Item Number/ AMT Description): X11-IS1 BUP-DX/ "Alere iScreen" Dx Single Dip Card Single Drug Detection: Buprenorphine (BUP)", X11-IS1 BAR-DX /"Alere iScreen" Dx Single Dip Card Single Drug Detection: Barbiturate (BAR)", X11-IS1 OXY-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Oxycodone (OXY)", X11-IS1 BZO-DX / "Alere iScreen" Dx Single Dip Card Single Drug Detection: Benzodiazepine (BZO300)", X11-IS1 MET-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Methamphetamine (MET)", X11-IS1 AMP-DX /"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Amphetamine (AMP)", X11-IS1 MDM-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Ecstasy (MDMA)", X11-IS1 MTD-DX /"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Methadone (MTD)", X11-IS1 MOP-DX/"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Morphine (MOP)", X11-IS1 THC-DX/"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Marijuana (THC)", X11-IS1 TCA-DX/ "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Tricyclic Antidrepessants (TCA)", X11-IS1 PCP-DX /"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Phencyclidine (PCP)",
Reason for recall
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.