Data Foundry

Medical device recalls 2018

VITROS Chemistry Products Cl- Slides — Class III medical device recall Z-3043-2018

Terminated recall by Ortho-Clinical Diagnostics, reported 2018-09-19, distributed nationwide. The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when usin

Recall numberZ-3043-2018
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2018-07-11
Classified2018-09-13
Reported by FDA2018-09-19
Terminated2020-08-11
DistributedNationwide (US)
Countries outside the USAU, BE, BM, BR, CA, CL, CN, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PL, PT, RU, SE, SG
Quantity128,196

Product

VITROS Chemistry Products Cl- Slides, Product Code 844 5207 Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum and plasma using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Reason for recall

The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the Cl- Slide was observed around a triglyceride concentration 600mg/dL, however the current triglyceride interference claim states interference should not occur below 800mg/dL of triglyceride.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3043-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.