QuickTox Drug Screen DipCard — Class II medical device recall Z-3043-2020
Terminated recall by Ameditech Inc, reported 2020-10-07, distributed nationwide. Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all ca
| Recall number | Z-3043-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ameditech Inc, San Diego, CA, United States |
| Recall initiated | 2020-08-18 |
| Classified | 2020-09-30 |
| Reported by FDA | 2020-10-07 |
| Terminated | 2024-09-30 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, DE, ES, FR, NZ, SG |
| Quantity | 170,425 units |
| Reason classes | labeling |
Product
QuickTox Drug Screen DipCard(AMT Item/ AMT Description): QT70/ "QuickTox¿ Drug Screen Dipcard 11 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+OXY100+MDMA500", QT88/ "QuickTox¿ Drug Screen Dipcard 12 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+OXY100+ MDMA500+BUP10", QT11 / "QuickTox¿ Drug Screen Dipcard 5 Drug Detection: COC300+OPI2000+THC50+AMP1000+PCP25", QT52/ "QuickTox¿ Drug Screen Dipcard 9 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+MTD300+OXY100", QT44/ "QuickTox¿ Drug Screen Dipcard 8 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+OXY100", QT23/ "QuickTox¿ Drug Screen Dipcard 6 Drug Detection: COC300+OPI2000+MET500+THC50+AMP1000+PCP25", QT80F/ "QuickTox¿ Drug Screen Dipcard 12 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+TCA1000+OXY100+MDMA500 with Positive Control Line", QT52A/ "QuickTox¿ Drug Screen Dipcard 9 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+MTD300+OXY100 4 Adulteration Test: Cr, Ni, pH, Ox ", QT20/ "QuickTox" Drug Screen Dipcard 5 Drug Detection: COC300+OPI2000+MET500+THC50+AMP1000"
Reason for recall
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.