VITROS Chemistry Products Cl- Slides — Class III medical device recall Z-3044-2018
Terminated recall by Ortho-Clinical Diagnostics, reported 2018-09-19, distributed nationwide. The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when usin
| Recall number | Z-3044-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class III |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ortho-Clinical Diagnostics, Rochester, NY, United States |
| Recall initiated | 2018-07-11 |
| Classified | 2018-09-13 |
| Reported by FDA | 2018-09-19 |
| Terminated | 2020-08-11 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, BE, BM, BR, CA, CL, CN, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PL, PT, RU, SE, SG |
| Quantity | 420,105 |
Product
VITROS Chemistry Products Cl- Slides, Product Code 684 4471 Product Usage For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum, plasma and urine using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. Product Usage:
Reason for recall
The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the Cl- Slide was observed around a triglyceride concentration 600mg/dL, however the current triglyceride interference claim states interference should not occur below 800mg/dL of triglyceride.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3044-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.