VITEK¿ 2 Compact 30 — Class II medical device recall Z-3076-2017
Terminated recall by Biomerieux Inc, reported 2017-09-06. Customers have reported that some VITEK¿ 2 cards are staying in preliminary status, not finalizing after eject
| Recall number | Z-3076-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Biomerieux Inc, Hazelwood, MO, United States |
| Recall initiated | 2017-07-12 |
| Classified | 2017-08-25 |
| Reported by FDA | 2017-09-06 |
| Terminated | 2020-12-16 |
| Quantity | 9,234 units |
| Reason classes | software |
| Model numbers | 27530, 27530R |
Product
VITEK¿ 2 Compact 30, REF numbers: 27530 and 27530R (clinical), and 27630 and 17630R (industry) software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.
Reason for recall
Customers have reported that some VITEK¿ 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK¿ 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.