Data Foundry

Medical device recalls 2017

CS 100i Intra-Aortic Balloon Pump — Class I medical device recall Z-3085-2017

Ongoing recall by Maquet Datascope Corp - Cardiac Assist Division, reported 2017-09-13, distributed nationwide. False blood back detection alarm and the ingress of fluids into the IABP affecting various electronic circuit

Recall numberZ-3085-2017
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMaquet Datascope Corp - Cardiac Assist Division, Mahwah, NJ, United States
Recall initiated2017-07-17
Classified2017-09-05
Reported by FDA2017-09-13
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KE, KR, KW, LB, LK, LT, LU, LV, MA, MT, MX, MY, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA

Product

CS 100i Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by augmenting blood pressure during the diastolic phase of the cardiac cycle. This increase in aortic pressure promotes more blood flow through the coronary arteries. Left ventricular work is reduced by decreasing aortic end diastolic pressure and reducing resistance to ventricular ejection, resulting in a decrease in blood pressure during the systolic phase of the cardiac cycle.

Reason for recall

False blood back detection alarm and the ingress of fluids into the IABP affecting various electronic circuit boards could potentially prevent initiation or continuation of therapy.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-3085-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.